Product features:
Measuring method: Boric acid affinity chromatography
Sample type: whole blood
Immediate testing: the samples are inspected as soon as they arrive.
Smart and efficient: It only takes 2 minutes and a half from sample addition to test completion.
Easy to operate: The calibration curve changes with the reagent, so you don't need to make your own standard curve.
Built-in timer: The instrument has a built-in timer, which can complete the test without additional timer.
Small and flexible: The palm-shaped instrument is convenient for field operations and areas with limited space.
Display: 2.4 inch color LCD touch screen, small and convenient
Built-in battery: Not affected by external power supply, can be used anytime, anywhere.
Device expansion: Support external Bluetooth printer, can print the report through Bluetooth printer
Application Department
Widely used in social health, outpatient, emergency departments of various hospitals, ICU, laboratory, endocrinology and other departments
Product parameters:
1 Results reporting unit
HbA1c is reported in IFCC (mmol/mol) and derived NGSP units (%).
2 Accuracy:
The reference substance is used as a sample for testing, and the relative deviation of the measurement results of the analyzer should be within ±8%.
3 Repeatability:
For samples with a concentration of 4.0%-6.5% (20.2mmol/mol-47.5mmol/mol), the coefficient of variation CV of the repeated measurement results of the analyzer should not be greater than 3.0%.
4 Linear:
In the detection range of the analyzer [4%-14%], the linear correlation coefficient r of the detection result should not be less than 0.9900.
5 Stability:
Within 8 hours after the start-up is stable, the relative deviation of the results of testing the same normal sample should not exceed ±3.0%.